WITE32 Release Notes Version 3.20
5
2
1
3
Figure 24: Old Servo Calibration Dialog Box
4
3
5
Figure 25: New Servo Calibration Dialog Box
4.2.4 Cool Down Delay in Servo Calibration
A cool down delay is added between the band erase operation and the measurements in the Servo Calibration
operation if the Erase Track Before Calibration option is enabled. Without this timeout, a continuous write
operation performed during the band erase overheats the head amplifier and the head cartridge. The thermal
expansion of a cartridge influences the measurements and might impair the calibration especially in the case of
narrow heads with small W/R offset.
50 Guzik Technical Enterprises